| Absorptance  The dissipation of light (radiation) within a surface or medium, caused 
      by the conversion of radiant (luminous) energy to a different form of energy, 
      usually heat, by interaction with matter. The absorption is the "missing piece", 
      when comparing the total reflected and transmitted energy with the incident 
      energy. The ratio of the total absorbed radiant or luminous flux to the incident 
      flux is called absorptance. The standard unit of absorptance is percent (%) 
      or a factor between 0 an 1. Absorptance can also be inferred from the transmission 
      through a medium. If the %transmission of a certain wavelength is 70% then 
      the material has a 30% Absorptance. Actinic  The characteristic of radiation that indicates its capability to produce 
      chemical change. In our industry the term is usually used with reference to 
      UV radiation and its effects on biological systems. Actinic strips are used 
      in UV processing to monitor the intensity of sources. The color or optical 
      density of the strip changes with the exposure dosage. Ambient Light Ambient light is the light diffused in the environment surrounding a detector 
      measuring the optical radiation from another source. This light contributes 
      to the signal measured from the source. To make valid measurements, the contribution 
      from the ambient light or background must be subtracted from each measurement.
       Aperture A hole through which radiant energy can pass. Angular aperture is the angle 
      through which the most divergent rays can pass through a hole or lens. The 
      aperture of a lens is often expressed in terms of an f/#. The f/# is the ratio 
      of the focal length of the lens to its diameter. A lens with a focal length 
      of 100mm and a diameter of 25mm would have an aperture of f/4. Attenuator A device which reduces the amount of energy reaching a sensor. Attenuators 
      are typically used when the radiant energy would saturate a detector. The 
      QNDS,QNDS2 and QNDS3 filters are attenuators that reduce flux density at the 
      detector by factors of 10, 100 and 1000, respectively. Top Bandwidth Bandwidth describes the size of a spectral segment. A bandwidth of 10nm 
      indicates a range of 10nm of radiation. This can be, for example, between 
      500nm and 510nm, 1000nm and 1010nm or an equal size segment anywhere in the 
      spectrum.  Band Eliminator Filter A band eliminator filter allows the wavelengths above and below the filter 
      cutoff to pass while suppressing the wavelengths within the band. These filters 
      are also called notch filters. A 500nm band eliminator filter with a band 
      pass of 10 nm would suppress the wavelengths between 495nm and 505nm. Beamwidth The angular width of a cone of light whose apex is at the source. Beamwidth 
      is usually defined as the angle subtended by a cone encompassing 90% of the 
      energy. Black Body A black body is an object that absorbs all radiant energy that hits it. 
      When heated a black body emits a well defined characteristic spectrum which 
      can be used in the characterizing the spectral responsivity of detectors. 
      As there is no such thing as a perfect black body, black body simulators are 
      used for this purpose. Top Calibration The process of normalizing the signal output from a detector to that of 
      a detector defined as a standard (usually defined by National Institute of 
      Standards and Technology (NIST) under identical illumination conditions). 
      Calibration can also be accomplished by the use of a standard source (lamp) 
      whose output energy at specific wavelengths and measurement distances are 
      traceable to the standard lamp defined by the governing standards body (NIST). Candela (cd) photometric measurement The SI unit of luminous intensity. One candela is equivalent to 1lumen/steradian. Candle photometric measurement 1candle = 1.02 candela (cd) Candlepower (cp) photometric measurement The luminous intensity of a source expressed in candelas. A source with 
      a luminous intensity of 10 candelas would be said to be a 10candlepower source.
       CCD A CCD (charge coupled device) is a light detector with high sensitivity 
      primarily in the visible spectrum. CCDs are typically made in either linear 
      or two-dimensional arrays consisting of up to millions of individual detector 
      elements. The 2D versions are used for image recording and are found in most 
      digital camera used in both scientific and consumer applications.  Chromaticity The aspects of color associated with hue and saturation without reference 
      to brightness. Chromaticity (CIE) Coordinates The proportion of the standard tristimulus values used in color matching. 
      Colors are compared by their CIE X, Y, and Z coordinates. CIE The CIE (Committee Internationale de l’Eclairage) is the international 
      standards organization for illumination and color vision. Color Temperature  Color temperature refers to the temperature, in degrees Kelvin, that a 
      black body would have to be heated to in order to have a color similar to 
      the reference. A 40W incandescent bulb has a color temperature of about 2680K 
      while daylight at noon has a color temperature of about 5500K. Cosine Collector A cosine collector is a translucent light collector that compensates for 
      normal blocking of radiation from flat surfaces. The cosine collector samples 
      the radiation according to the cosine law through the hemisphere above the 
      surface. A cosine collector may also be referred to as a Lambertian surface. Cosine Law (Lambert’s cosine law) The flux per unit solid angle leaving or entering a surface is proportional 
      to the cosine of the angle with respect to the normal to the surface. In a 
      cosine collector the rays that strike the surface at an angle of 60° from 
      the vertical will have a contribution of 0.5 (cosine of 60°) that of identical 
      rays arriving vertically. Cutoff Filter A filter that does not pass light of wavelengths shorter that the cutoff 
      wavelength and passes a wide band of wavelengths above the cutoff wavelength. 
      The cutoff wavelength is specified at some point along the transition from 
      maximum transmission to 0 transmission. The same filter can have different 
      cutoff wavelengths depending upon the specified % transmission. The same filter 
      may be specified as a 500nm cutoff filter with the 50% transmission point 
      as the reference or a 485nm cutoff filter if the specification is the 5% transmission 
      point, where the transmission at 485nm is 5%. Top Dark Adaptation The ability of the human eye to adjust itself to low light levels. Dark Signal (Current) The dark signal is the signal that flows in a photodetector when there 
      is no optical radiation impinging upon it. This signal is created within the 
      detector and the amplification circuits due to thermionic (temperature) effects. Densitometer A densitometer measures the opacity or absorptance of a material. The measurement 
      is usually expressed in AU (absorptance units) or OD (optical density). Diffraction Grating A diffraction grating is an optical component that separates light into 
      its constituent wavelengths. Functionally equivalent to a prism, it disperses 
      the light into its spectrum by employing grooves to diffract the light. The 
      angle of the diffraction is a function of the wavelength. Diffuse Reflectance The ratio of the incident flux to the reflected flux from a scattering 
      surface as opposed to a highly directional or specular (mirror-like) surface. Dose A common, but loosely used, term for energy density, or radiant flux density, 
      at a surface. (It is a precisely defined term in EB curing: 1 Gray (Gy) = 
      1 J/kg , a measure of absorbed energy per unit mass). In other technologies, 
      the term usually applies to energy absorbed within the medium of interest, 
      but in UV curing, is equated only to irradiant energy density arriving at 
      the surface of the medium of interest. [The preferred shortened term is energy 
      density, expressed in J/cm² or mJ/cm²]. Dynamic Range The dynamic range is ratio of the maximum measurable signal before saturation 
      to the minimum measurable signal above the noise. Typically the dynamic range 
      is expressed in either decades (powers of 10) or bits (power of 2). A 5 decade 
      dynamic range indicates that there is a factor of 100,000 between the maximum 
      and minimum signals measurable with the apparatus. A dynamic range of 16 bits 
      (264) indicates a factor 65,532 between the minimum and maximum measurable 
      signals.  Dynamic range is also expressed in db (decibels), defined as 10 log10 (maximum 
      signal / minimum signal). Five decades of dynamic range is equal to 50 db. Top Einstein A unit of energy equivalent to the amount of energy absorbed by one molecule 
      of a material undergoing a photochemical reaction as determined by the Stark-Einstein 
      law. Electromagnetic Radiation Radiation emitted by vibrating charged particles. A combined oscillation 
      of electric and magnetic fields that propagates through space at the speed 
      of light. The electromagnetic spectrum is theoretically infinite, includes 
      gamma, x-ray, UV, visible, IR, microwaves, and radio waves. Emissivity The ratio of an object’s radiance to that emitted by a black body at the 
      same temperature and wavelength. Energy Density Radiant energy arriving at a surface per unit area, usually expressed in 
      joules or millijoules per square centimeter (J/cm2 or mJ/cm2). It is the time-integral 
      of irradiance. (Other terms applied include "radiant exposure," "light dose," 
      and "total effective dosage"). Etendue Also called the throughput of an optical system, it is the product of its 
      entrance aperture and the solid angle in which light can be accepted at that 
      aperture. Exitance Flux leaving a surface per unit area. Top Flux The energy per second (power) in a light beam expressed in Watts or Joules/second. 
      (Radiant power). In photometric measurements the luminous power is typical 
      expressed in lumens (lm) . Footcandle (fc) photometric measurement Footcandles is a unit of measurement of illuminance (how much light is 
      striking a point on a surface). Once footcandle is equivalent to 10.764 lux. Footlambert (fl) photometric measurement A unit of luminance equal to 1/p candela /ft2. Top Germicidal All biological organisms contain DNA. DNA is essential to reproduction. 
      Optical radiation in the UVC range is capable of breaking the molecular bonds 
      within DNA effectively killing microorganisms. Germicidal UV lamps are used 
      for water treatment, sterilization of foods and their containers, and air 
      purification, especially in hospital environments. Hefner Unit photometric measurement  A unit of luminous intensity equal to 0.9 candle. Illuminance photometric property Luminous flux incident per unit area of a surface. 1 lumen/m2 = 1lux. Top Infrared (IR) The invisible portion of the electromagnetic spectrum that extends from 
      0.75 microns to 1000 microns. Radiation in the near infrared (NIR) produces 
      the sensation of heat. Integrating Sphere A hollow sphere coated with a white diffusing coating in the interior. 
      It is used for measuring the diffuse reflectance and transmittance of objects 
      or the total flux from a source that is completely inside. Intensity Flux per solid angle. Radiometric measurements are made in W/sr. Photopic 
      measurements are made in lumens/sv. Inverse Square Law The inverse square law correlates the relative intensity at varying distances 
      from a point source. The relative intensity will diminish to a factor of the 
      square root of the difference in distance. For example if at 2 meters from 
      a source the intensity is 16 W/m2, it will be 4W/m2 at 4meters and W/m2 at 
      8 meters. For extended (non-point) sources the intensity fall off approaches 
      the inverse square law at a distance equivalent to 5 times the diameter of 
      the source. Irradiance radiometric property Radiant flux incident per unit area of a surface; the power incident per 
      unit area. The radiometric unit of measure is W/m2 or factors thereof (mW/cm2). 
      The photometric units of measure are lumens/m2, lux, phot, and footcandles. Top Joule (J) The Joule is the SI unit of energy.  Lambertian Surface A surface whose emission or scattering follows Lambert's Cosine Law, in 
      which the radiant intensity leaving a surface is proportional to the cosine 
      of the angle from surface normal. See Cosine Collector. Linearity The precision with which there is a direct relationship between the incident 
      radiation and the resultant measurement value up to a point of saturation. 
      A linearity of 1% states that the ratio of the measured value to that of the 
      incident radiation will not vary more than 1% from the absolute. Lumen (lm) photometric measurement The lumen is the photometric unit of power. It is the flux emitted in a 
      unit solid angle by a point source that has one candela luminous intensity. Luminance photometric property Flux density per unit solid angle. Lux radiometric measurement S.I. unit of illuminance equal to 1 lumen per square meter. Top Micron A unit of length equal to 10-6 m. Infrared wavelengths are typically measured 
      in microns.  Monochromator A monochromator is a devise that uses a diffraction grating or prism to 
      disperse light into a spectrum of its constituent wavelengths. The dispersive 
      element is rotated such that only a narrow (monochromatic) band of light is 
      permitted to exit the monochromator through a narrow aperture or slit. Nano  Prefix denoting 10-9. One nanowatt (nW) = 10-9 watts. Nanometer (nm) A unit of length equal to 10-9 m.  Abbreviated nm. Commonly used unit 
      to define wavelength of light, particularly in the UV and visible ranges of 
      the electromagnetic spectrum.  Narrow Band Filter A narrow band filter allows only a limited number of wavelengths to pass 
      through it. Narrow band filters are usually specified at a specific central 
      wavelength, a band pass indicating the range of wavelengths that will pass 
      through it and a % transmission at the limits of the band pass. A 500nm narrow 
      band pass filter with a band pass of 10 nm with a 5% cut off will pass the 
      wavelengths between 495nm and 505nm. The transmittance above and below these 
      wavelengths will be less than 5%.  Neutral Density Filter A filter that reduces the intensity of the light passing through it without 
      altering the relative spectral distribution of the energy. Neutral densities 
      are given by the log base 10 of their attenuation. An attenuation of 100 gives 
      a Neutral Desnity (ND) of 2. See Optical Density. Nit (nt) photometric measurement A unit of measure of brightness (luminance) equal to one candela per square 
      meter. Noise Equivalent Irradiance (NEI) The radiant flux density in W/cm2 required to produce a signal equal to 
      the inherent noise of the detection system. The input irradiance at which 
      the signal to noise ratio is 1. Noise Equivalent Power (NEP) The radiant power, at a specified wavelength and band pass that will produce 
      a an output signal from a detector that is equivalent to the inherent noise 
      in that detector. Normal  The normal is the axis drawn perpendicular to an illuminated surface. The 
      normal is the reference from which reflective, diffractive and refractive 
      angles are measured. A ray with a zero degree angle of incidence arrives perpendicular 
      to a surface. A ray with a 90° angle of incidence is parallel to a surface 
      and may not strike it. Notch Filter See band eliminator filter. Top Opacity A measure of a material's ability to block light. It is equivalent to the 
      reciprocal of the material’s transmittance. Optical Chopper A mechanical or electro optical device for passing and interrupting, at 
      a uniform frequency, a beam of light. Optical Density (OD) A measure of the transmittance T through an optical medium.  OD = 
      -log10T.   An OD of 1 is equivalent to 10% transmission. A 2 OD 
      filter would have a transmission of 1%. Peak Irradiance UV curing The intense peak of focussed power directly under a lamp. The maximum point 
      of the irradiance profile. Measured in irradiance units (W/cm²). Phot (ph) photometric measurement A measurement unit for illuminance. One phot = 10,000 lux (lx). Photodiode A photodiode is a two electrode semiconductor device with an optical radiation 
      sensitive junction in which the reverse current varies with the illumination. 
      The wavelength sensitivity is a function of the materials used in the device. 
      Silicon photodiodes are sensitive through most of the visible spectrum. InGaAs 
      photodiodes are sensitive in the NIR region of the spectrum. GaP photodiodes 
      are used for the UV region of the spectrum. Photodynamic Therapy The use of optical radiation in the cure of medical maladies. Photodynamic 
      therapy is used in the treatment of skin conditions such as psoriasis, jaundice 
      in newborns, and more recently in the treatment of certain types of cancer. Photometer  A device for measuring luminous intensity or luminance. A photometer employs 
      a photopic filter which has a band pass that is matched to the human eye response. 
      The S.I. units employed are lumens and lux.  Photomultiplier Tube (PMT) A photomultiplier tube is a vacuum device in which a photocathode emits 
      electrons when exposed to light. The electrons are then accelerated through 
      electrostatic fields to metal plates where a greater number of secondary electrons 
      are emitted. This is repeated through several stages. This amplifies the current 
      by many thousands. Photopic  Having sensitivity characteristic similar to the human eye response. A 
      photopic filter will have a band pass between 400 and 700nm with its highest 
      transmission at 550nm with a spectral responsivity specified by CIE. Photoresist A chemical substance which is rendered insoluble upon exposure to light. 
      By exposing photoresist through a mask, electrical circuits can be created 
      by washing the unexposed areas and etching the material below. Photoresists 
      are usually optimized for specific regions of the UV spectrum specified as 
      UVC, UVB, and UVA depending on the type of lamp employed for the exposure. Photostability Many chemical products both medicinal and protective coatings can degrade 
      when exposed to light. Photostability measurements are made to determine the 
      principal wavelengths responsible for the degradation and the amount of exposure 
      (dose) required to create a change that will be deleterious to the product’s 
      efficacy. In photostability studies the effects from the visible wavelengths 
      of light must be distinguished from the UV. Typically two detectors are employed 
      with filtration limiting each to the measurement to only one spectral region.
       Pico (p) Prefix denoting 10-12. One pW = 10-12 Watts. Top Radiance radiometric measurement Radiant power per unit source area per solid angle. W/m2/steradian. Radiant Emittance radiometric measurement Radiant power emitted into a full sphere (4p steradian) by a unit area 
      of a source , expressed in W/m2. An Integrating sphere is typically used to 
      make this measurement. Radiant Exitance radiometric measurement The radiant flux per unit area emitted from a source. Radiometer A device for measuring the intensity or accumulation of radiant energy. Radiometry The science of radiation measurement. The detection and measurement of 
      radiant energy either at specific wavelengths or band passes or as a function 
      of wavelength over a broad spectrum. The measurement of the interaction of 
      light with matter as to absorption, transmission, and reflectance. Ray The geometric representation of a light path through an optical system. Reflectance The ratio of the reflected flux to the incident flux from a surface. In 
      some cases the measurement may be made with either the specular or diffuse 
      component of the total reflected flux. Reflectance is expressed as a percent. Relative Spatial Responsivity The relative spatial responsivity of a detector indicates the acceptance 
      angle and percent of incident radiation at that angle that will strike the 
      detector. The measurement is usually made in comparison to a perfect Lambertian 
      surface. Responsivity (spectral sensitivity) The response or sensitivity of any system as a function of incident wavelength. 
      In radiometry, it is the output of a device versus wavelength. Top Saturation A condition wherein the radiant flux density exceeds either or the capacity 
      of a photodetector to emit electrons in a linear relationship to the incident 
      flux and/or the current produced by the detector exceeds the capacity of the 
      electronics to measure the current in a linear fashion. Scotopic  Relating to the wavelength responsivity of the human eye under dark adapted 
      conditions. Sensitivity The ratio of the output signal from a detector to the input signal. This 
      may also be expressed as the minimum input irradiance level which will produce 
      an output signal that exceeds the noise level of the detector; i.e. where 
      the S/N is greater than 1. SI Systeme Internationale d’Unities; the international metric system of units.
       Spectral Response The measure of a detector’s relative sensitivity as a function of incident 
      wavelength. A typical spectral response curve will display the responsivity 
      as a percentage at a given wavelength to wavelength of maximum responsivity. Spectrometer / Spectrograph  A device which measures the interaction between light and materials as 
      a function of wavelength. A spectrometer is usually a monochromator with an 
      integrated detector. A spectrograph does not have an exit slit, allowing a 
      broad band of wavelengths to be measured simultaneously by means of a multi 
      element detector or photographic plate. Specular Reflectance Reflection from a mirror-like surface where in the coherence of the incident 
      beam is maintained in the reflected beam. This is opposed to diffuse reflection 
      wherein the reflected light is spread in all directions in a Lambertian-like 
      manner.  Steradian (sr) The unit of solid angle subtended by an area on the surface of a sphere 
      equal the square of the radius of the sphere. One steradian can be visualized 
      as a conic section with a solid angle of approximately 66°. Stilb (sb) photometric measurement A unit of luminance equal to 1candela/cm2. Top Talbot photometric measurement The SI unit of the quantity of light expressed in lumen-seconds. Thermocouple Thermopile A device made up of dissimilar metals in which a small current is produced 
      as a function of the difference in temperature of the materials at the junction. 
      Thermocouples can be used for measurement of radiation in the infrared region 
      of the spectrum. Transmittance The ratio of the radiant power transmitted through a material to the incident 
      radiant power. Transmittance is usually expressed as a percent. A filter with 
      a 50% transmittance (at a specific wavelength) will absorb half of the light 
      incident on it and allow half of it to pass through it. UV (ultraviolet) The invisible portion of the electromagnetic spectrum with wavelengths 
      between 1nm and 400nm. UVA The portion of the UV spectrum covering the wavelength range between 320nm 
      and 400nm. This spectral region is used for many medical, UV curing and photolithographic 
      applications. The earth’s atmosphere (at sea level) absorbs all wavelengths 
      shorter than UVA. Prolonged exposure to UVA radiation will cause sunburn. UVB UVB is the portion of the UV spectrum that covers the wavelength range 
      between 280nm and 320nm. UVB radiation is typically used in UV curing and 
      photolithographic applications as well as in certain medical applications. 
      Exposure to UVB radiation (from lamps or electric arcs) can cause severe sunburn 
      and cause eye damage. UVC UVC is the portion of the UV spectrum that extends from 190nm to 280nm. 
      UVC is usually employed in water treatment and sterilization applications. 
      UVC is also used in UV curing and photolithography in microelectronics applications. 
      Exposure to UVC radiation (from lamps, arcs, or lasers) can cause severe biological 
      damage. Top VUV (vacuum ultraviolet) The VUV is the portion of the UV spectrum below 190 nm. Electromagnetic 
      radiation below 190nm is absorbed by oxygen in air. Physical or chemical interactions 
      requiring VUV radiation must be performed in a nitrogen purged environment 
      down to 160nm or in a vacuum chamber below 160nm.  Visible Spectrum (VIS) The visible portion of the spectrum extends between 400nm and 700nm (per 
      the CIE). It covers the wavelengths of light that the human eye can perceive. Watt (W) radiometric measurement The Watt is a unit of power or work. One Watt corresponds to one Joule/sec. Wavelength When electrons vibrate they produce oscillating perpendicular electric 
      and magnetic fields. The distance between sequential field intensity maxima 
      is defined as the wavelength. These distances for visible radiation are very 
      small and are usually expressed in unit lengths of nanometers (nm).  Top |